Particle counters are
monitoring to provide evidence of controlled performance related to air
cleanliness by particle concentration (ISO-14644-2) according to
“Classification of air cleanliness by particle concentration” (ISO-14644-1). In
Medico Support we can provide the calibration of particle counters in
compliance to ISO-21501-4 for correct determination of particle size
distribution.
The ISO-21501-4 is
describing the following parameters:
-
Size
setting error;
-
Counting
efficiency;
-
Size
resolution;
-
False
count;
-
Maximum
particle number concentration;
-
Sampling
flow rate error;
-
Sampling
time error;
-
Response
rate;
-
Calibration
interval;
-
Reporting
results from test and calibration
Beside the parameters
described in 21501-4, there are a number of other parameters, which is relevant
to monitor. Like Threshold setting voltage, Laser diode voltage and Stray
light. The extra monitoring can give valuable information about the status of
equipment and life-cycle.
The calibration can be
performed accredited, in principle calibration is carried out in our own
laboratories, but in some cases it is also possible to calibrate at the
customer.
The calibrations are
performed by our experienced and trained staff who are always up to date with
the latest knowledge and know the latest requirements. They can also provide
advice when purchasing new measuring equipment.
If you have any questions or just want to hear more, contact Medico Support.
For some equipments we can also make a total service agreement, which include all services, spare parts, exchange of batteries, repair, adjustments and recalibration.