- Mærke: Michell Instruments
Tags: OptiCal, Humidity, Calibrator, Michell, Instruments
The OptiCal works by time proportioning the flows of dry and saturated air according to the pre-selected relative humidity value. The chamber temperature is controlled by a four-zone Peltier heat pump arrangement for maximum stability and minimum temperature gradient. The saturator assembly can be filled easily with the distilled water reservoir at the front of the unit and the easily accessible desiccator can be re-charged by heating when required. No other maintenance is necessary. The only external service required is a single phase power supply.
Product Features
- Precision relative humidity and temperature calibrator
- Generate 10 to 90% RH over +10 to +50°C (+50 to +122°F) temperature
- Integral chilled mirror reference hygrometer
- UKAS accredited calibration, as standard
- Humidity and temperature profile generation for unattended verification of sensors
- Transportable to allow on-site calibrations
Technical Specifications
| Humidity | |
| Generated Range | 10 - 90% RH |
| Control Element Accuracy | ≤ ±1% RH (10-70% RH) ≤ ±1.5% RH (70-90% RH) |
| Stability | ±0.2% RH (20-80% RH) |
| Temperature | |
| Generated Range | +10 to +50°C (+50 to +122°F) |
| Accuracy | ±0.1°C (±0.18°F) |
| Accuracy | ±0.1°C (±0.18°F) |
| Chamber | |
| Ramp Rate | +20 to +40°C (+68 to +104°F): 1.5°C / minute (2.7°F / minute)
+40 to +20°C (+104 to +68°F): 0.7°C / minute (1.2°F / minute) |
| Control Element | Removable relative humidity sensor |
| Reference | |
| Accuracy | Dew Point: ±0.2°C (±0.36°F) Temperature: ±0.1°C (±0.18°F) |
| Measurement Units | Dew Point: (°C/°F), % RH Temperature: (°C/°F), gm-3, gkg-1, water activity (aW) |
| Outputs | Analog: 4-20 mA or 0-20 mA over user-settable output Accuracy: ±0.2°C (±3.6°F) Digital: 500 Ω maximum load resistance Alarm: RS232 @ 9600 baud rate Volt free contact, 30 V, 100 mA maximum |
| General | |
| Probe Ports | Up to 5 - sensor body diameters 5 to 25mm (0.2 to 0.98") accommodated by port adapters |
| Chamber Volume | 2000cm3 (112.1in3) |
| Chamber Dimensions | 105 x 105 x 160mm (4.13 x 4.13 x 6.3") (h x w x d) |
| Instrument Dimensions | 290 x 520 x 420mm (11.4 x 20.5 x 16.5") (h x w x d) |
| Set-point Resolution | 0.1 for humidity and temperature |
| Displays | 2 line Vacuum Fluorescent Display |
| Supply | 85 to 264 V AC, 47/63 Hz, 150 V A |
| Weight | 20kg (44lbs) |
Applications