- Brand: Michell Instruments
Tags: XZR400, Series, Oxygen, Analyzers, Michell, Instruments
A range of trace oxygen analyzers utilizing Michell's Metallic Sealed Reference Sensor (MSRS)
technology. Designed to measure O2 as an impurity in pure inert gases such as N2, CO2,, He, Ar
and Kr. The XZR400 can also be used for cylinder filling and other applications where a precise
measurement of oxygen is required. There are 4 different chassis to choose from and various
internal options including pumps, flow alarms and digital communications to match customer
needs.
See datasheet for the full specification of the XZR400 Series range.
Product Features
- Fast response time
- Simple and easy operation
- Low maintenance and cost of ownership
- No need for reference air
- Barometric pressure compensation
- Multiple Outputs available from 4-20 mA, RS485 Modbus RTU and RS232
- Rack, wall-mount, bench-mount, and transportable versions available
- Optional built-in Easidew Transmitter with rack-mount version
Sensor Type | |||||
Measurement Principle | Zirconium Oxide Sensor with Metallic Sealed Reference and S Type thermocouple | ||||
Performance | |||||
Gas | Clean, dry, oil free with particles less than 3μm | ||||
Measurement Range | 0.1 ppm up to 25% O2 Extended ranges available up to 0 to 100% O2 | ||||
Lowest detectable limit (LDL) | 0.1ppm(v) O2 | ||||
Accuracy (Intrinsic Error) | Less than 2% of reading | ||||
Response Time | <11 seconds | ||||
Repeatability | ±0.1% of reading | ||||
Stability | 1% per month | ||||
Linearity | Better than ±1% | ||||
Drift | <1% of reading per week | ||||
Sample Flow Rate | 1 to 3 Nl/h with built-in fast loop | ||||
Maximum Sample Pressure | 2 barg(29 psig) | ||||
Maximum Sample Temperature | +100°C (+212°F) | ||||
Atmospheric Pressure Compensation | Built-in as standard | ||||
Optional sensor | |||||
Moisture sensor | Easidew sensor can be fitted as an option. Please, see a separate datasheet for full specifications. | ||||
Measurement range | -100 °C to + 20 °C | ||||
Sample Flow rate | 60 to 300 Nl/hour | ||||
Outputs | |||||
Analog Output | 0, 1 or 2 off 0/4-20 mA Linear with Galvanic Isolation output | ||||
Digital Communications Output | Modbus RTU over RS485 protocol (Standard on XZR400A1) RS232 output (optional) |
||||
Output Load | Over 1000 Ω | ||||
Self-diagnostics | Via HMI | ||||
Output Ranges | 0.0 ppm to 1ppm up to 0 to 25% O2 Extended ranges available up to 0 to 100% O2 |
||||
Alarms | 2 threshold alarms, freely configurable 1 general fault alarm including flow alarm 1 flowalarm (optional) |
||||
Display Resolution | 0.01 ppm between 0.1 ppm and 10 ppm 0.1 ppm between 10 ppm and 10,000 ppm 0.01% between 1 to 10% 0.1% between 10 to 25% |
||||
Power Supply | 90 to 264 V AC, 47/63Hz | ||||
Power Consumption | 50 VA | ||||
Operating Conditions | |||||
Ambient Temperature Range | 0 to +55°C (+32 to +131°F) | ||||
Sensor Temperature | Optimized @ +634°C | ||||
Operating Humidity | 5 to 90% RH without condensation | ||||
Mechanical Specification | |||||
Model | Dimensions | Weight | Gas connectors | Ingress Protection | |
Rack mount: | 19’’, 3U, 482.5 x 133 x 371.5mm | 10kg | 1/8" | IP20 | |
Wall mount: | 200 x 220 x 290mm | 5kg | 6 mm | IP40 | |
Bench mount: | 290 x 260 x 236mm | 5.2kg | 6 mm | IP20 | |
Transportable: | 450 x 300 x 330mm | 13.5kg | 6 mm | IP40 | |
Sample Gas Connections | All gas connections are Swagelok (MALE) Fittings (316 SS) |
|