- Brand: Michell Instruments
Tags: MDM300, I.S., High, Speed, Dewpoint, Hygrometer, Portable, Hygrometers, Michell, Instruments
A high-speed portable dew-point hygrometer, offering rapid spot-check measurements of dew point or moisture content in many applications, including compressed air, natural gas and high voltage switchgear quench gas. This lightweight, ATEX, IECEx, FM, CSA, GOST and INMETRO certified product allows more measurements per working hour than any other comparable product. A hard-wearing but ergonomic case and an easy-to-use interface allows comfortable operation in the toughest industrial environments.
Product Features
- Repeatedly fast measurements at low pressure in less than 15 minutes for T95 down to -60°C dew point (-76°F dew point)
- Higher pressure measurements possible – up to 35 MPa (350 barg / 5076 psig)
- Long battery life: up to 48 hours of typical use between charges
- Intuitive application kits allow quick and straightforward connection to your sample point
- Durable, yet easy to handle and operate: designed for use in industrial environments
- 4-20mA external device input for transmitter calibration and validation
- Lightweight: less than 1.5kg (3.31 lbs)
- 13 point traceable calibration certificate
Technical Specification | MDM300 | MDM300 Intrinsically Safe |
Performance | ||
Measurement Technology | Michell ceramic sensor | |
Accuracy | ±1°C from -60 to +20°C (±1.8°F from -76 to +68°F) dew point ±2°C from -100 to -60°C (±3.6°F from -148 to -76°F) dew point ±0.2°C temperature (0.36°F) | |
Calibrated Range | Spot Checks: -70 to +20°C (-94 to +68°F) dew point Online Analysis: -100 to -70°C (-148 to -94°F) dew point | |
Uncalibrated Readings From | +20 to +30°C (+68 to +86°F) dew point | |
Measurement Units | °C, °F, K dew point & gas temperature ppmV, ppmW for air, N2, H2, CO2, SF6 %RH, g/m3, g/kg Option: active pressure (bara/g, psig, MPa, KPa) | °C, °F, K dew point & gas temperature ppmw & g/kg for air, N2, H2, CO2, SF6 ppmv, lb/MMscf & g/m3 for natural gas ppmv, g/m3 & %RH |
Resolution (Display) | 0.1 for all dew-point derived units and autoranging where appropriate | |
Resolution (Measurement) | Better than 0.1°C (0.18°F) dew point | |
Typical Response Speed | T95 in ≤15 minutes to -60°C (-76°F) dew point | T95 in ≤30 minutes to -60°C (-76°F) dew point |
Measurement System Repeatability | Better than 0.1°C (0.18°F) dew point | |
Measurement System Hysteresis | ±0.05°C (0.09°F) dew point | |
Electrical Input/Output | ||
Auxiliary Inputs | 4-20 mA loop-powered external input selectable as either dew point, temperature or pressure | Connection to Michell Easidew TX I.S. or Easidew PRO I.S. via Remote Sensor Interface |
Battery Type | NiMH 4.8V | |
Battery Operating Life | Up to 48 hours of typical usage between charges | Up to 24 hours of typical usage between charges |
Battery Charger | Intelligent charger (supplied) | Intelligent charger (not certified for hazardous area use) |
Operating Conditions | ||
Operating Pressure | 35 MPa (350 barg / 5076 psig) max (Please consult your Michell Sales Engineer for advice on operating at pressure exceeding 15 MPa (150 barg / 2175 psig) | |
Operating Environment | Outdoors 0 to +100% RH condensing | |
Operating Temperature | -20 to +50°C (-4 to +122°F) | |
Storage/Transport Temperature | -40 to +70°C (-40 to +158°F) | |
Mechanical Specification | ||
Display | Blue LCD graphical display | |
Enclosure Type | Steel fiber-loaded high-impact polyamide 6 | |
IP/NEMA Rating | IP66/NEMA 4 | |
Gas Connections | 1/8” NPT female (other options available) | |
Flow Across Sensor | 0.2 to 1.2 Nl/min (0.42 to 2.54 scfh) | 0.2 to 0.5 Nl/min (0.42 to 1.05 scfh) |
Gas Wetted Materials | AISI 316L stainless steel | AISI 316L stainless steel, PTFE Seal, Borosilicate glass, ceramic |
Outline Dimensions | 218 x 170 x 90mm (8.6 x 6.7 x 3.5") (d x w x h) | |
Weight | 1.35kg (2.98lbs) | 1.5kg (3.31lbs) |
General | ||
Data Logging | 8 megabytes; Log interval: 5 to 60 sec; Logs per log file: Up to 10,000 | |
Communications | (Wireless) BluetoothTM range up to 5m (16.4’) (version 2.0) | |
User Interface Languages | English, French, German, Italian, Portuguese, Spanish | |
Certification Codes | ||
| CE | ATEX & IECEx – EX Group II Cat 1G Ex ia IIC T4 Ga -20°C to 60°C CSA/FM – NEC 500 Class1 Div 1 Groups A,B,C,D T4 INMETRO: Ex ia IIC T4 Ga (-20°C to +50°C) GOST-R, GOST-K CE |
|