- Brand: Michell Instruments
Tags: Trace, Oxygen, Analyzer, for, Process, Industries, Michell, Instruments
Trace Oxygen Analyzer for Process Industries
GPR-1800 IS, GPR-1800 AIS
The GPR-1800 series of intrinsically safe trace oxygen analyzers are designed for use in demanding process environments where the low detection limit down the AII’s industrial oxygen sensors can be well utilized. Typical for Analytical Industries gas analyzers, they are simple to use as well as easy and cost-effective to maintain thanks to the use of their maintenance-free galvanic oxygen sensors. The innovative liquid drain manifold, which is available as an option, protects and extends the sensor's life span in processes where liquids may be present in the sample gas.
Product Features
- Four measurement ranges provided
- LDL of 50 ppb
- Measures in CO2 with XLT sensor
- Sensor life of up to 24 months
- Simple, intuitive HMI
- Barometric pressure and temperature compensation
- Two user configurable alarms
- Optional Modbus
- Range of sampling options available
Technical Specifications
GPR-1800 IS | GPR-1800 AIS | |||
Measurement Range | 0-10, 0-100, 0-1000 ppm, 0-1%, 0-25% (for calibration only) |
|||
Accuracy | < 2% of selected range at constant conditions | |||
Response time | T90 < 10 seconds T90 < 2 minutes (for Liquid drain models) | |||
Sensitivity (LDL) | 0.05ppm | |||
Linearity (LDL) | <1% of scale | |||
Sensor model | GPR-12-333 or GPR-12-333-LD XLT-12-333 or XLT-12-333-LD for gases containing > 0.5% CO2 |
|||
Sensor life at 25ºC (77°F) and 1 atm | 24 months in < 1000 ppm O2 5°C (77°F)and 1 atm | |||
Calibration interval | 30 days | |||
Inlet pressure | 0.34-2 barg (5-30 psig) with atmospheric vent | |||
Flow rate | 0.5-1.0 l/m (1-2 SCFH) | |||
Gas connections | 1/8" compression tube fittings 1/4" compression tube fittings (with liquid drain option) |
|||
Wetted Parts | Stainless steel | |||
Display | Graphical LCD 7 x 3.5cm (2.75 x 1.375"); resolution 0.01 ppm | |||
Enclosure | UL: NEMA Type 3R ATEX: NEMA 4x |
|||
Compensation | Temperature | Barometric pressure and temperature | ||
Signal output | 4-20mA Loop current | 4-20 mA or 1-5V | ||
Communications | NA | Optional Modbus RTU (AIS 24 V version only) | ||
Alarms | N/A | Two user configurable alarms: magnetic coil relays rated 3A at 100 VAC |
||
Operating temperature | GPR sensor: 5°C to 45°C (41°F to 113°F) XLT sensor: -10° to 45°C (14°F to 113°F) |
|||
Power | 18-24 V DC 2wire loop power | 12-28 V DC or 110-220 V AC | ||
Liquid Drain | Liquid Drain option available for both models | |||
Approvals | CULUS: Class I, Division 1, Groups C & D Tamb -20° to +50°C ATEX: II 2 G Ex d [ib] ib IIB T4 Gb |