Pressure Swing dryers provide a source of very dry compressed
air for use as a zero gas in humidity calibration systems, or for general laboratory applications.
The Michell PSD Series Pressure Swing Dryers use two columns
filled with 4Å molecular sieve desiccant,
which are used alternately on a two-minute switching cycle. The
PSD dryers are designed to operate continuously,
using a small proportion of the dried air, to regenerate the
offline column - generally giving desiccant life in excess of
five years. This type of heatless regeneration also uses
significantly less energy than a 'heated' type of dryer.
Product Features
- Excellent long term stability
- Maintenance free except for a desiccant change once every 5 years
- Completely self-contained
- Low power consumption
| Model | PSD2 | PSD4 (Standard) | PSD4-HFV (High flow volume) | PSD4-HPO (High pressure output) |
| Performance |
| Gas output |
| Flow | 7 Nl/min (14.8 scfh) | 30 Nl/min (63.6 scfh) | 90 Nl/min (109.7 scfh) | 30 Nl/min (63.6 scfh) |
| Pressure | 0.5 barg (7 psig) | User-settable up 8 barg (116 psig) |
| Moisture content | <1ppmv | <13.8ppbv |
| Input Requirements |
| Gas supply |
| Flow | 10 Nl/min (21.2 scfh) | 30 Nl/min (63.6 scfh) | 90 Nl/min (109.7 scfh) | 30 Nl/min (63.6 scfh) |
| Pressure | 5 to 7 barg (70 to 100 psig) | 6 to 10 barg (87 to 145 psig) |
| Moisture content | Oil and liquid water-free | <16ppmv |
| Electrical Input |
| Power | 100 to 115 OR 220 to 240 VAC, 50/60Hz | 100 to 240VAC 50/60Hz |
| Power connection | IEC socket |
| Environmental Conditions |
| Operating temperature | +5 to +35°C (+41 to +95°F) | +10 to +40°C (+50 to +104°F) |
| Storage temperature | -40 to +35°C (-40 to +95°F) | -40 to +50°C (-40 to +122°F) |
| Mechanical Specifications |
| Type | Twin column desiccant, pressure swing |
| Desiccant | 4 Ångström Molecular sieve bead (4-8 mesh) |
| Timer | Mechanical cam | Programmable relay |
| Gas connections |
| Inlet | Swagelog ® ¼" | ¼" VCR Swagelok® |
| Outlet | Swagelog ® ¼" | ¼" VCR Swagelok® | ½" VCR Swagelok® | ¼" VCR Swagelok® |
| Fliters |
| Outlet | None |
Millipore Wafergard IIF Micro Inline (sealed type) with PTFE membrane
element rated at >99.999% retention of 0.003μm particles |
| Vent | None | Bonded glass microfiber rated at >99.999% retention of 0.1μm particles |
| Construction | Rack mount: 19" x 6U x 324mm (12.8") | 304 stainless steel wall mounting enclosure: 800x600x300mm (31.5x24x12”) |
| Weight | 12.5kg (27.5lbs) | 30Kg (66lb) | 35Kg (77lb) | 32Kg (70lb) |
Applications |
 |
Related Products |
Calibration Instruments |
 |
DCS The DCS system is a premium customizable rack-mounted dew-point calibration station
with cooled mirror reference. |
|
 |
Dew-Point Generators A range of dew-point generators offering ease of use and fast response to humidity step changes. |
|
Cooled Mirror Reference Instruments |
 |
Optidew
Vision The Optidew Vision precision dew-point meter is based on the
proven, fundamental optical dew... |
|
 |
S8000 Remote A hygrometer offering a combination of a market-leading ±0.1°C accuracy with the flexibility of a remote sensor.. |
|
 |
S8000
Features an integrated sensor head allowing measurement
in the range of -60°Cdp to +40°Cdp, combining the measurement
sensitivity and ±0.1°C dew point accuracy of the RRS system. |
|
 |
S8000 RS The
S8000 RS chilled-mirror sensor directly measures the formation of
condensation, giving long-term and unmatched, drift-free readings of
dew-point and relative humidity. |
|
 |
S4000 TRS The S8000 TRS chilled-mirror sensor directly measures the formation of condensation, giving long-term and unmatched... |
|
| |