Primary Flow Calibration Standard
For High Capacity Gas Flow
Featuring Proven DryCal® Technology, the DryCal 1020 and 1500 provide high-capacity primary gas flow calibration from 5 to 1500 standard liters per minute (slpm). The compact piston-prover system provides a practical alternative to traditional large bell provers for laboratory and calibration applications.
DryCal Technology provides accurate and reliable gas flow readings independent of the gas species measured. For use with non-corrosive, non-condensing, non-combustible gases at less than 70% humidity.
Documentation
Specifications, manuals and calibration procedures from Mesa Labs.
Choose the Right Model
DryCal 1020
- Flow range: 5–500 slpm
- Standardized accuracy: 0.25% of reading
- Operating pressure: 10–19.5 PSI absolute
DryCal 1500
- Flow range: 15–1500 slpm
- Standardized accuracy: 0.45% of reading
- Operating pressure: 45 PSI absolute
- Upstream flow measurement capacity: 1500 slpm
System Specifications & Features
- Total flow range: 5–1500 slpm
- Gas sensor input factor
- Accurate flow measurement independent of gas species measured
- Upstream flow measuring capacity up to 1500 slpm
- Dimensions: 24" W × 30" H × 12" D
- Weight: 90 lb
- 1/2" inlet and outlet Swagelok® compression fittings
- Serial RS-232 data port
- Plug-and-play installation with simple push-button operation
- Compact alternative to traditional large bell provers
MFC & MFM Calibration
DryCal primary piston provers can be combined with Mesa Labs Integrator Pro to create an analog MFC and MFM calibration station. Integrator Pro can power and control up to two voltage and two current mass flow devices simultaneously.