FlexCal® Series
Highly Accurate Portable Gas Flow Calibrators
The Mesa Labs FlexCal® Series combines standardized and volumetric primary flow measurement with high accuracy, portability and simple operation. With three models covering flow ranges from 5 ccm to 50 LPM, FlexCal is designed for demanding laboratory, industrial hygiene and field calibration applications.
All FlexCal models provide 0.5% of reading accuracy and measure temperature and pressure directly in the flow stream. User-selectable volumetric and standardized flow units, a graphical LCD display and USB connectivity make setup and data collection straightforward.
For use with non-corrosive, non-condensing, non-combustible gases at less than 70% humidity.
Documentation
Current technical documentation and manufacturer information from Mesa Labs.
FlexCal Models
| Model | Flow Range | Accuracy | Temperature / Pressure Sensor |
|---|---|---|---|
| FlexCal-L | 5–500 ccm | 0.5% of reading | Yes, in flow stream |
| FlexCal-M | 50 ccm–5 LPM | 0.5% of reading | Yes, in flow stream |
| FlexCal-H | 0.5–50 LPM | 0.5% of reading | Yes, in flow stream |
Measurement & Operation
- Volumetric units: mL/min, L/min, cc/min, cf/min
- Standardized units: smL/min, sL/min, scc/min, scf/min
- Temperature: °C / °F
- Pressure: mmHg / PSI / kPa
- Integrated temperature and pressure sensors
- Pressure or suction flow modes
- Approximate measurement time: 1–15 seconds
- Single, Continuous and Burst measurement modes
- User-selectable averaging from 1 to 100 measurements
- Backlit graphical LCD display
- USB data interface
- Compatible with DryCal Pro software
Portable by Design
- Dimensions: 170 × 159 × 73.5 mm
- Weight: 1.36 kg
- Operating pressure: atmospheric to 15 PSIA maximum
- Storage temperature: 0–70°C
- 6V rechargeable sealed lead-acid battery
- 6–8 hours typical battery operation
- Low / Medium: 1/4" ID fittings
- High: 3/8" ID fittings
- CE / RoHS compliant
Accessories
Mesa Labs offers accessories for FlexCal and other DryCal flow calibrators, including power and charging accessories, tubing, adapters, leak-test accessories and protective carrying solutions.
View DryCal accessories ↗Calibration & Traceability
Mesa Labs recommends annual verification and calibration of the FlexCal by an ISO/IEC 17025 accredited laboratory or authorized service provider to maintain measurement integrity. Recertification can include inspection, testing, available upgrades and a NIST-traceable calibration certificate.
Calibration & service information ↗